Tantalum balloon-expandable stent: in vivo swine studies. Journal of vascular and interventional radiology Fontaine, A. B., Dake, M. D., TSCHANG, T. P., Guertin, S., Stabbe, M. T., Dos Passos, S. 1993; 4 (6): 749-752

Abstract

The authors describe the experimental use of a balloon-expandable tantalum vascular stent in normal and atherosclerotic microswine.Thirty-one stents (15 iliac, 13 aorta, two femoral, one renal) were placed in 11 animals. Stents were placed in both normal and stenotic atherosclerotic arteries. The animals were killed and the stents explanted 2-32 weeks after placement. Arteriography was performed just prior to death in all cases.All stents were patent arteriographically and on gross inspection. Histopathologic study showed a thin, smooth, endothelial-lined neointima over the segments with stents. This covering had a mean thickness of 2.9 x 10(-5) mm and appeared maximally developed 2-5 weeks after stent placement. The stent design incorporates the following engineering and clinical considerations: minimal profile and surface area and maximal expansion ratio, hoop strength, flexibility, fluoroscopic visibility, biocompatibility, and ease of placement. When compressed and mounted on the balloon catheter, the stent is flexible. This allows it to be easily maneuvered through tortuous or diseased vessels. On expansion, the stent becomes a semirigid tubular structure composed of multiple rhomboid cells. Stents with an expanded diameter of less than 11 mm can be introduced through a 7-F vascular sheath.Initial results suggest that this stent successfully incorporates a number of theoretically desirable features for a vascular prosthesis, and it may offer some advantages over previously described devices.

View details for PubMedID 8280995