Failure of Fracture Plate Fixation JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS Gardner, M. J., Evans, J. M., Dunbar, R. P. 2009; 17 (10): 647-657

Abstract

Failure of fracture fixation after plating often leads to challenging surgical revision situations. Careful analysis of all patient and fracture variables is helpful in both determining the causes of the fixation failure and maximizing the success of subsequent interventions. Biologic and mechanical factors must be considered. Biologic considerations include traumatic soft-tissue injury and atrophic fracture site. Common mechanical reasons for failure include malreduction, inadequate plate length or strength, and excessive or insufficient construct stiffness. Reliance on laterally based implants in the presence of medial comminution may be a cause of fixation failure and subsequent deformity, particularly with conventional nonlocking implants. Management of dead space with cement or beads has been effective in conjunction with staged approaches. An antibiotic cement rod in the diaphysis may provide fracture stabilization. Locking full-length constructs should be considered for osteoporotic fractures.

View details for Web of Science ID 000270475300007

View details for PubMedID 19794222