Improved uniformity of aortic enhancement with customized contrast medium injection protocols at CT angiography RADIOLOGY Fleischmann, D., Rubin, G. D., Bankier, A. A., Hittmair, K. 2000; 214 (2): 363-371

Abstract

To compare the uniformity of aortoiliac opacification obtained from uniphasic contrast medium injections versus individualized biphasic injections at computed tomographic (CT) angiography.Thirty-two patients with an abdominal aortic aneurysm underwent CT angiography. In 16 patients (group 1), 120 mL of contrast material was administered at a flow rate of 4 mL/sec. In the other 16 patients (group 2), biphasic injection protocols were computed by using mathematic deconvolution of each patient's time-attenuation response to a standardized test injection. Attenuation uniformity was quantified as the "plateau deviation" of enhancement values, which were calculated as the SD of the time-contiguous attenuation values observed during the 30-second scanning period.Group 2 patients received between 77 and 165 mL (mean, 115 mL) of contrast medium. Initial flow rates ranged from 4.1 to 10.0 mL/sec (mean, 6.8 mL/sec) for the first 4-6 seconds; continuing flow rates ranged from 2.0 to 4.8 mL/sec (mean, 3.1 mL/sec) for the remaining 24-26 seconds. The plateau deviation was significantly smaller in group 2 patients (19 HU) versus group 1 patients (38 HU, P <.001).At CT angiography, tailored biphasic injections led to more uniform aortoiliac enhancement, compared with standard uniphasic injections of contrast medium.